- Physical & Mechanical Tests
Cell Ageing AS/NZS 1660

The cell oven has a stirrer that stirs the air in the cells with about 60 rpm, which meets the standard AS/NZS 1660.2.3:1998. This cell oven is equipped with 4 cells, each with individual temperature. Each cell has individual control of the air exchange rate for precision ageing of rubber and plastic materials under controlled conditions.
- No. of Cells: 4/6/8 – As per customer requirement
- Inner Chamber made of Aluminium as described in relevant specification.
- Direct Air Flow Measurement for 4-20 air changes using digital air flow meters (made in JAPAN)
- Temperature Controller: Digital PID Controller
- TOUCH SCREEN INTERFACE
- Temperature Range: RT+40 ~ 150 Deg. C
- Suitable to perform ageing test for insulation and sheathing material at different temperature ranges e.g., 80 Deg. C, 100 Deg. C, 135 Deg. C and 150 Deg. C
- Automatic measurement of air replacement rate 3-20 air changes per hour
- Temperature Accuracy: +/- 1 Deg C
- Digital Timer with cut off
- Over Temperature Limiter
- Stirrer 60 RPM (adjustable)
- Confirms to AS/NZS 1660.2.3
- Temperature Range 40 Deg.C to 150 Deg.C (Test Temperature)